Dual Channel K-Type Thermocouple Thermometer
HI935002
Description
The Dual Channel K-Type Thermocouple Thermometer – HI935002 is a digital, waterproof thermometer. This meter offers high accuracy temperature measurements using a wide variety of interchangeable probes for the two available input channels.
•  °C/°F readout
•  High and low-temperature display
•  Two channels
Features at-a-glance :
°C/°F Readout
•  Measurements can be displayed in either degrees Celsius or Fahrenheit. With a simple switch on the back of the meter, the meter can display readings in the °C or °F scale.
  Battery Error Prevention System (BEPS)
•  The Battery Error Prevention System detects when the batteries become too weak to ensure reliable measurements.
High/Low Function
•  The maximum and minimum temperature values are continuously monitored and displayed on the lower portion of the HI935002 LCD display during a measurement session.
•  The CLR button clears the high and low values on the LCD display.
HOLD Function
•  The HOLD button on the face of the meter freezes the display to allow the user time to record readings.
•  Although the display is frozen, the meter continues to internally monitor the temperature and update the high and low measurement values.
Interchangeable Probes
•  A wide range of K-type thermocouple probes are available to meet the specific needs of users.
•  Any of the HI766 series of probes can be interchanged with the HI935002 to measure the temperature of surfaces, gases, air, liquid, semi-solid samples, and more.
Low Battery Indicator
•  When the battery level is below 10%, a warning symbol will blink to indicate low battery condition.
  Two Channels
•  HI935002 can monitor samples through two independent temperature channels.
•  With two probes connected to the thermometer, temperature can be monitored in two samples at the same time.
•  With both channels in use, a press of the T1/T2 button allows users to display the desired channel and measurement
Two Display Modes
•  The LCD can display standard or relative measurement mode. In standard mode, the display shows the current temperature and the high and low limits for the appropriate channel.
•  In relative mode, the current temperature is set as the reference temperature for the appropriate channel; the display then shows the difference from the reference temperature.
  SPECIFICATIONS
| Range (°C) : | -50.0 to 199.9°C and 200 to 1350°C | 
|---|---|
| Range (°F) : | -58.0 to 399.9°F and 400 to 2462°F | 
| Resolution (°C) : | 0.1°C (-50.0 to 199.9°C) and 1°C (outside) | 
| Resolution (°F) : | 0.1°F (-58.0 to 399.9°F) and 1°F (outside) | 
| Temperature Accuracy (@25ºC/77ºF) : | ±0.2% FS (excluding probe error) | 
| Electrode/Probe : | HI766 series K-type thermocouple (not included) | 
| CAL Button : | yes | 
| Backlit LCD : | yes | 
| Battery Type/Life : | 1.5V AA (3) / approximately 1600 hours of continuous use | 
| Auto-off : | 8 min (default), 60 min or OFF | 
| Environment : | -10 to 50°C (14 to 122°F); RH max 100% | 
| Dimensions : | 150 x 80 x 36 mm (5.9 x 3.1 x 1.4’’) | 
| Weight : | 235 g (8.3 oz) | 
| Ordering Information : | HI935002 is supplied with batteries and instructions. | 
Order now