Description
Control of Standard Components
| ControlCOCKPIT | adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays |
|---|---|
| Temperature | 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value |
| Timer | Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Temperature
| Set temperature range in °C | min. 5°C above ambient up to +300°C |
|---|---|
| resolution of display for setpoint values 0.1°C up to 99.9°C, 0.5°C from 100°C and for actual values 0.1°C (LED) | |
Control Technology
| adjustable parameters | temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime |
|---|---|
| Function HeatBALANCE | adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 % |
| Function SetpointWAIT | the process time does not start until the set temperature is reached |
| Language setting | German/English/Spanish/French |
| Calibration | three freely selectable temperature values |
Ventilation
| natural convection | |
|---|---|
| Fresh air admixture | adjustment of pre-heated fresh air admixture by air flap control in 10 % steps for each segment individually |
| Vent | vent connection with restrictor flap |
Communication
| Programming | AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
|---|---|
| Documentation | programme stored in case of power failure |
Safety
| Alarm | audible and visual |
|---|---|
| Temperature control | overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display |
| AutoSAFETY | additionally integrated over- and undertemperature monitor “ASF”, automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature |
| Temperature control | mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
| Autodiagnostic system | for fault analysis |
Standard Equipment
| Door | fully insulated stainless steel door with 2-point locking (compression door lock) |
|---|---|
| Internals | 2 stainless steel grids |
| Scope of delivery | incl. works calibration certificate for +160°C |
| Housing | rear zinc-plated steel |
| Interior | easy-to-clean interior,made of stainless steel,reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides |
Stainless Steel Interior
| w(A) x h(B) x d(C): 640 x 800 x 500 mm | |
| Volume 256 l | |
| Max. loading of chamber: 300 kg |
Textured Stainless Steel Casing
| w(D) x h(E) x d(F): 824 x 1183 x 684 mm |
Electrical Data
| 230 V, 50/60 Hz / approx. 3400 W | |
| 115 V, 50/60 Hz / approx. 2200 W |
Packing/Shipping Data
| the appliances must be transported upright | |
|---|---|
| Customs tariff number | 8419 8998 |
| Country of origin | Federal Republic of Germany |
| WEEE-Reg.-No. | DE 66812464 |
| Dimensions approx incl. carton B x H x T: 930 x 1380 x 930 mm | |
| Net weight approx. 110 kg | |
| Gross weight carton approx. 161 kg | |